Senzor de monitorizare atmosferică Winson ZEHS04 Manual de utilizare al modulului
Declaraţie
This manual copyright belongs to Zhengzhou Winsen Electronics Technology Co., LTD. Without the written permission, any part of this manual shall not be copied, translated, stored in data base or retrieval system, also can’t spread through electronic, copying, record ways. Thanks for purchasing our product. In order to let customers use it better and reduce the faults caused by misuse, please read the manual carefully and operate it correctly in accordance with the instructions. If users disobey the terms or remove, disassemble, change the components inside of the sensor, we shall not be responsible for the loss. The specific such as color, appearance, sizes …etc., please in kind prevail. We are devoting ourselves to products development and technical innovation, so we reserve the right to improve the products without notice. Please confirm it is the valid version before using this manual. At the same time, users’ comments on optimized using way are welcome. Please keep the manual properly, in order to get help if you have questions during the usage in the future.
ZEHS04
Profile

ZEHS04 este un modul multi-în-un de tip difuzie, montat cu modulul de monitorizare atmosferică ZE12A, pentru a detecta CO, SO2, NO2 și O3. De asemenea, este compatibil cu modulul senzor de praf, modulul senzor de temperatură și umiditate extern. Cu ieșire TTL sau RS485, este convenabil de utilizat și de depanat, ceea ce scurtează foarte mult ciclul de proiectare și dezvoltare al utilizatorului și răspunde nevoilor clienților pentru diferite ocazii de detectare a gazelor.
Caracteristică
Sensibilitate ridicată, rezoluție înaltă, durată lungă de viață;
Ieșire UART sau RS485;
Stabilitate ridicată, capacitate bună anti-interferență, ieșire liniară excelentă;
Aplicație
Monitorizarea mediului atmosferic urban;
Monitorizarea emisiilor neorganizate de poluare la amplasamentele fabricii;
Instrumente portabile, echipamente de monitorizare a calității aerului și echipamente pentru casă inteligentă.
Caietul de sarcini

Interval de detectare

Protocolul de comunicare
1. setari generale
Tabelul 3

2. Comenzi de comunicare
Setările implicite sunt modul de încărcare inițiativă. Modulele încarcă valoarea concentrației de gaz la fiecare 1 S,
Tabelul 4

Notă: convertiți hexazecimal în zecimal înainte de calcul;
Valoarea concentrației gazului = Gaz (octet mare)*256+ Gaz (octet scăzut)
Valoarea temperaturii= (Temp. high byte*256+ temp. low byte – 500)*0.1
Valoarea umidității= (Umiditate. octet înalt*256+ temperatură. octet scăzut)*0.1
Dacă se adaugă funcția de pompare, pompa este activată implicit. Formatul de comandă pentru oprirea pompei este următorul:
Stabil 5.

Pentru a deschide funcția de pompare: Stabil6.

Sumă de control și calcul
caracter nesemnat FucCheckSum (caracter nesemnat *i, caracter nesemnat ln)
{
caracter nesemnat j,tempq=0;
i+=1;
pentru(j=0;j<(ln-2);j++)
{
tempq+=*i;
i++;
}
tempq=(~tempq)+1;
return(tempq);
}
Sugestie Shell:
- The peripheral structure must be water-proof. The front and back sides of casing, need to be opened to ensure that the air can diffuse freely for testing.
- The module is provided with a fixing hole that it can be fixed to the outer casing through the fixing hole.
- If it’s pumping type, there should be a hole with 3mm or more diameter, on the casing, to facilitate the airpipe to draw out the outside air.
Atenționări:
- Vă rugăm să nu utilizați modulele în sisteme care au legătură cu siguranța omului.
- Vă rugăm să nu expuneți modulele în gaz organic cu concentrație mare pentru o perioadă lungă de timp.
- Senzorul trebuie să evite solvenții organici, acoperirile, medicamentele, uleiul și gazele cu concentrație mare.
- Modulul ar trebui să fie încărcat mai mult de 24 de ore pentru prima dată, iar circuitul de alimentare trebuie să fie echipat cu funcție de rezervare a energiei. În caz contrar, va afecta continuitatea și acuratețea datelor returnate dacă acestea sunt offline pentru prea mult timp. Dacă timpul de oprire a alimentării este în mai puțin de o jumătate de oră, trebuie să fie învechit timp de cel puțin 2 ore.
- Se recomandă menținerea îmbătrânirii senzorului și oprirea pompei pentru a economisi energie, precum și a prelungi durata de viață a pompei și a asigura acuratețea datelor senzorului, atunci când modulul nu este testat.
- Conform protocoalelor de comunicare, este necesar să se verifice dacă byte0, byte1 și valoarea de verificare sunt corecte după primirea datelor, astfel pentru a asigura corectitudinea recepționării cadrelor de date.
- Se sugerează să utilizați instrumentele USB – convertire – TTL și software-ul asistent de depanare UART și să observați pe baza protocoalelor de comunicare pentru a aprecia dacă comunicarea cu modulul este normală.
Zhengzhou Winsen Electronics Technology Co., Ltd.
Adăuga: No.299, Jinsuo Road, National Hi-Tech Zone, Zhengzhou 450001 China
Tel: + 86-371-67169097 / 67169670
Fax: +86-371-60932988
E-mail: sales@winsensor.com
Website-ul: www.winsen-sensor.com
Documente/Resurse
![]() |
Modulul senzor de monitorizare atmosferică Winson ZEHS04 [pdfManual de utilizare ZEHS04 Modul senzor de monitorizare atmosferică, ZEHS04, Modul senzor de monitorizare atmosferică, Modul senzor de monitorizare, Modul senzor atmosferic, Modul senzor, Modul Senzor, Senzor |