MEXTECH DT1503

MEXTECH DT1503 Digital Capacitance Meter User Manual

Model: DT1503 (MEX105)

1. Introducere

Mulțumesc pentru achizițieasing the MEXTECH DT1503 Digital Capacitance Meter. This precision instrument is designed for accurate capacitance measurements across a wide range, suitable for laboratory, factory, and educational use. Please read this manual thoroughly before operating the device to ensure safe and correct usage.

Continutul pachetului:

MEXTECH DT1503 Digital Capacitance Meter package contents including meter, test leads, 9V battery, and user manual.
Figure 1: MEXTECH DT1503 package contents.

The image above shows the MEXTECH DT1503 Digital Capacitance Meter along with its included accessories: a set of test leads, a 9V battery, and the user manual.

2. Informații de siguranță

WARNING: To avoid electrical shock, do not operate the meter with the battery cover open.

CAUTION: This device complies with IEC61010-1 standards, rated for 1000V CAT II and 600V CAT III, Pollution Degree 2. Always adhere to safety guidelines.

Spate view of the MEXTECH DT1503 Digital Capacitance Meter showing safety warnings and battery compartment.
Figura 2: Spate view of the meter with safety markings.

The image displays the back of the MEXTECH DT1503 meter, highlighting important safety warnings and compliance information printed on the battery compartment cover.

3. Caracteristicile produsului

The MEXTECH DT1503 Digital Capacitance Meter offers the following key features:

Faţă view of the MEXTECH DT1503 Digital Capacitance Meter showing its display, rotary switch, and buttons.
Figura 3: Față view of the MEXTECH DT1503 meter.

Această imagine oferă o imagine clară view of the meter's front panel, including the LCD display, rotary function switch, 'LIGHT' and 'HOLD' buttons, and the 'ADJ' knob for zero adjustment.

4. Configurare

4.1 Instalarea bateriei

  1. Ensure the meter is turned OFF by rotating the function switch to the 'OFF' position.
  2. Localizați compartimentul bateriilor pe spatele contorului.
  3. Open the battery cover by sliding or unscrewing it (refer to Figure 2).
  4. Insert one 9V battery, observing the correct polarity (+ and - terminals).
  5. Închideți bine capacul bateriei.

4.2 Conectarea cablurilor de testare

  1. Insert the red test lead plug into the 'CX+' input terminal.
  2. Insert the black test lead plug into the 'COM' input terminal.
Faţă view of MEXTECH DT1503 Digital Capacitance Meter with red and black test leads connected to the CX+ and COM terminals.
Figure 4: Test leads connected to the meter.

This image shows the MEXTECH DT1503 meter with the red test lead connected to the 'CX+' terminal and the black test lead connected to the 'COM' terminal, ready for measurement.

5. Instrucțiuni de utilizare

5.1 Pornire/Oprire

To turn the meter ON, rotate the rotary function switch from the 'OFF' position to any desired measurement range. To turn the meter OFF, rotate the switch back to the 'OFF' position.

5.2 Ajustare la zero

Before performing any capacitance measurement, especially on lower ranges, it is crucial to perform a zero adjustment to eliminate stray capacitance from the test leads and internal circuitry.

  1. Ensure the test leads are connected to the meter.
  2. Short the tips of the red and black test leads together.
  3. Rotate the 'ADJ' knob (located on the left side of the meter) until the display reads '000'.
  4. This adjustment should be performed for each range if significant deviation from zero is observed.

5.3 Măsurarea capacității

  1. Condensator de descărcare: Always ensure the capacitor to be measured is fully discharged before connecting it to the meter. Failure to do so can damage the meter or cause injury.
  2. Selectați intervalul: Rotate the function switch to the appropriate measurement range. If the capacitance value is unknown, start with the highest range (20mF) and work downwards until a stable and accurate reading is obtained.
  3. Condensator de conectare: Connect the test leads to the capacitor terminals. For polarized capacitors (e.g., electrolytic), connect the red lead to the positive terminal and the black lead to the negative terminal. For non-polarized capacitors, the connection order does not matter.
  4. Valoare citită: The capacitance value will be displayed on the LCD screen in the selected unit (pF, nF, µF, or mF).

5.4 Funcția de păstrare a datelor

To freeze the current reading on the display, press the 'HOLD' button. The 'HOLD' indicator will appear on the screen. Press the 'HOLD' button again to release the reading and resume live measurement.

5.5 Lumina de fundal

Press the 'LIGHT' button to activate the display backlight. This feature is useful for reading measurements in low-light conditions. The backlight will automatically turn off after approximately 10 seconds to conserve battery power.

6. Intervale de măsurare și precizie

The MEXTECH DT1503 offers the following measurement ranges and typical accuracy specifications:

GamăPrecizieRezoluţie
200pF±0.5% + 20 cifre0.1pF
2nF±0.5% + 20 cifre1pF
20nF±0.5% + 20 cifre10pF
200nF±0.5% + 20 cifre100pF
2 uF±0.5% + 20 cifre1nF
20 uF±0.5% + 20 cifre10nF
200 uF±0.5% + 20 cifre100nF
2mF±2% + 20 cifre1 uF
20mF±2% + 20 cifre10 uF

7. Întreținere

7.1 Curățare

To clean the meter, wipe the casing with a dry, soft cloth. Do not use abrasive cleaners, solvents, or harsh chemicals, as these may damage the plastic components or electronic circuits.

7.2 Înlocuirea bateriei

When the low battery indicator appears on the display, the 9V battery needs to be replaced. Follow the steps outlined in Section 4.1 'Battery Installation'. Ensure the meter is OFF before replacing the battery.

7.3 Înlocuirea siguranței

If the meter fails to function or provides incorrect readings consistently, the internal fuse may need replacement. Fuse replacement should only be performed by a qualified technician to ensure the correct fuse type and rating are used. Opening the meter by unauthorized personnel may void the warranty.

8. Depanare

This section addresses common issues you might encounter with your MEXTECH DT1503 meter.

9. Specificații

ParametruCaietul de sarcini
Afişa3½ Digit LCD, 1999 counts
Alimentare electricăBaterie de 9V (tip NEDA 1604 sau 6F22)
Temperatura de operare0°C până la 40°C (32°F până la 104°F)
Temperatura de depozitare-10°C până la 50°C (14°F până la 122°F)
Dimensiuni (L x l x H)185 x 90 x 53 mm
GreutateAproximativ 380 g (inclusiv bateria)
Numărul de modelMEX105 (also known as DT1503)
Evaluare de siguranțăIEC61010-1, CAT II 1000V, CAT III 600V, Pollution Degree 2

10. Garanție și asistență

For warranty information, technical assistance, or service inquiries regarding your MEXTECH DT1503 Digital Capacitance Meter, please refer to the warranty card included with your product or visit the official MEXTECH website pentru date de contact și resurse de asistență.

Documente conexe - DT1503

Preview MEXTECH Digital Dual Display TRMS Clamp Manual de utilizare a contorului
Comprehensive operation manual for the MEXTECH Digital Dual Display TRMS Clamp Meter (DT-3600), detailing safety precautions, component identification, technical specifications, and step-by-step operating instructions for various electrical measurements.
Preview Generator de funcții Mextech FG3006 - Generare de semnale de înaltă performanță
Explorați Mextech FG3006, un generator de semnal monocanal portabil, rentabil și multifuncțional. Dispune de tehnologie DDS, un afișaj color de 2.4 inci, ieșire sinusoidală de 5 MHz și o varietate de forme de undă. Include specificații, aplicații și accesorii.
Preview MEXTECH DT-369 Digital True RMS Clamp Manual de utilizare al contorului
User manual for the MEXTECH DT-369 Digital True RMS Clamp Meter, covering features, specifications, safety instructions, and operation.
Preview MEXTECH Digital Manometer MM45A - Pressure Measurement Tool
Detaliat pesteview of the MEXTECH Digital Manometer MM45A, featuring its applications, key features, technical specifications, and accessories for precise air and gas pressure measurement.
Preview Mextech Digital Manometer MM45B 5PSI - Technical Specifications and Features
Detailed information on the Mextech Digital Manometer MM45B, including its applications, features, and specifications for measuring air pressure in various systems.
Preview Manometer User Manual - MM45C Model - MEXTECH
User manual for the MEXTECH MM45C Manometer, detailing its features, specifications, product description, and operational instructions for measuring air and differential pressure.